Hearing Evaluation

Diagnostic Audiologic testing

These tests are part of a diagnostic test battery performed by Dr. Veronica Heide to determine the degree and type of your hearing ability. The outcome of these tests determines whether you need medical referral or audiologic habilitation, hearing instruments, use of assistive devices, speech reading lessons, cochlear implant referral, or other therapies.

As well as other appropriate tests of perception necessary to understand how your ears and hearing system are working.

How hearing works

Ear Anatomy.jpg

Outer Ear

Using video otoscopy, you and the audiologist can see if your outer ear canal is partially or fully occluded with earwax. If there is no debris blocking the ear canals, you will be able to clearly see the landmarks of the eardrum (tympanic membrane) and the middle ear bones.

Video Otoscopy

Video Otoscopy

Video Otoscopy of the eardrum

Video Otoscopy of the eardrum

Middle Ear

Middle Ear

Middle Ear

The middle ear system is an impedance matching device that helps transform the airborne sound waves efficiently from the outer ear through the middle ear to the inner ear. If there is a problem with the conduction of sound through the middle ear, sound energy will be lost and hearing loss may occur. Middle ear problems are most often medically managed and may include a hole in the eardrum, ear diseases such as middle ear fluid and infections, stiffening of the middle ear bones (otosclerosis) and other mechanical problems. If middle ear problems are detected during the evaluation you will most likely be referred to an otolaryngologist for medical evaluation and to be cleared medically before pursuing hearing devices.

Audiologists can measure how well the middle ear is working using Tympanometry and Acoustic Reflex measurements. Equipment has evolved from measuring the middle ear dynamics from a single frequency to measuring multiple frequency responses from one measurement of reflected sound in the ear canal.

Responses using a single tone measurement

Responses using a single tone measurement

 
Wide band tympanometry

Wide band tympanometry

MiddleEarResults.jpg
AR REflex and Decay.jpg

Inner Ear (Hearing and Balance)

Inner Ear

Inner Ear

 

The inner ear is the sensory organ for both hearing and balance. A hearing loss in the inner ear is most often called, “sensorineural loss,” but recently new testing has become available to allow audiologists to differentiate sensory loss from neural loss.

CochleaPiano.png
BrainMap.png

Otoacoustic Emission Testing

Otoacoustic emission testing is often used to evaluate the integrity of the outer hair cells in the inner ear. These microscopic structures are the pre-amplifier of the inner ear elongate and contract to optimize the frequency and intensity of the input signal. Newborn hearing screening has been made possible by the development of equipment to measure outer hair cell function.

Normal outer hair cell structure

Normal outer hair cell structure

Abnormal outer hair cell structures

Abnormal outer hair cell structures

 
Measurement of outer hair cell function in infant hearing screening

Measurement of outer hair cell function in infant hearing screening

 

The top graph shows a robust normal Transient Otoacoustic Emission (TOAE) from an infant, while the bottom graph shows a lack of response from a different infant. Infants are screened in the newborn nursery and referred to a pediatric audiologist for further diagnostic testing if the baby does not pass the screening. Audiologists may also use special computerized tests of the auditory nerve called Auditory Brainstem Response tests to determine if the loss of hearing is due to loss of sensory function or loss of transmission of the auditory signal from the cochlear through the auditory nerve and pathway to the brain.

For more information photographs and videos about ear anatomy go to: http://www.cochlea.eu/en/auditory-brain/brain-stem

Return to top of page

Hearing threshold assessment

Harvey Fletcher was a pioneer in audio measurement and sound transmission. His plot below, the Fletcher-Munson curves (later termed equal loudness contours) was derived from his research at Bell Labs to improved telephone audio fidelity. Through this work he pioneered the invention of the first hearing aid and invented stereo sound and demonstrated this at the Chicago World’s Fair much to the delight of those who attended.

Fletcher Munson curves.jpg
Harvey Fletcher.png

Fletcher, H. and Munson, W.A. "Loudness, its definition, measurement and calculation",
Journal of the Acoustical Society of America 5, 82–108 (1933).

The range of frequencies heard by the human ear is between 20 Hz to 20,000 Hz, but the ear is most sensitive in the frequency range from 2000 Hz through 5000 Hz due to resonance of the ear canals and the mechanical advantage provided by the ossicles.

Rather than plot hearing test results on a Fletcher-Munson curve, audiologists took the sound pressure level curve that defined average normal hearing across frequency and termed that average normal hearing with a reference of 0 dB hearing level (HL). So, instead of a curved line, the results are plotted on a straight line referenced to 0 dB HL.

Audiologists typically test the “speech frequencies” between 250 Hz through 8000 Hz in routine diagnostic testing but can extend the range of frequencies tested by using ultra-high frequency headphones calibrated with special audiometric equipment to test as low as 125 Hz and as high as 16,000 Hz.

Conventional Audiogram

Conventional Audiogram

Ultra-high frequency audiogram

Ultra-high frequency audiogram

Degree of Hearing Loss:

  • -10 to +15 dB – Normal hearing range

  • 15 to 25 dB – Slight loss – misses some soft and distant sounds

  • 30 to 40 dB loss – Mild loss – not aware of soft and distant sounds

  • 45-55 dB loss – Moderate loss – misses many critical speech sounds – often blames others for not speaking clearly.

  • 60-70 dB loss – Moderately severe loss. Misses much of speech without amplification. Generally pleased with hearing instruments.

  • 75 to 90 dB loss – Severe loss - Loss of OHCs and some IHCs results in lack of tuning ability and poor speech understanding even with the best hearing instruments. Speech Reading and the use of ALDs are keys to successful communication.

  • 95 to >110 - Profound loss – Need visual with auditory to communicate effectively. Will benefit from use of captioning, written text, and visual cues. May be a cochlear implant candidate.

Return to top of page

Perception

So, while audiologists have automatic tests of the functional integrity of the various components of the hearing mechanism, they still ask patients to respond to tones and speech to measure the perceptions of the individual. The ability to detect sounds across a variety of frequencies and intensities is plotted on an audiogram along with tests that measure speech detection and word comprehension in quiet and in noise.

Return to top of page

Speech Understanding

In addition to testing audibility for tones, audiologists will evaluate the softest level at which you can hear speech and also how clearly you understand speech in quiet and speech in noise. The ability to hear well in noise is based on many factors including cognition, neural processing and life experiences. The ability to understand speech in noise can be improved with training.

Unerstanding Speech in Noise.png

To learn more about the role of neural processing, cognition, and life experience, visit the website of Nina Kraus, Ph.D.: https://brainvolts.northwestern.edu/

We Hear With Our Brains.png

Loudness Discomfort Testing

Loudness discomfort testing will ensure that loud sounds sound loud but not uncomfortably so. This measurement defines the top end of your comfortable listening range and is important not only in hearing instrument programming, but in understanding sound tolerance problems.

Treatment Recommendations

At the conclusion of your diagnostic evaluation, your audiologist will know a lot about your ears and how you perceive sound. By putting the pieces of the diagnostic testing together the audiologist will understand the type of hearing loss and the degree of hearing loss and the impact of these factors on your lifestyle and communication. Based on the diagnostic testing, treatment recommendations are made. These recommendations may include:

  • Hearing instruments

  • Continued audiologic follow-up and programming

  • Assistive listening devices

  • Aural habilitation such as LACE listening program, music training or personal therapy

    Return to top of page